User guide
14 Electrical Connections
4.4 AC ELECTRICAL CONNECTIONS
Danger
Make sure that power to the instrument is disconnected
before making any wiring connections.
CONNEXIONS AUX COURANT SECTEUR / ALTERNATIF
Danger
Débrancher toute alimentation à l'appareil avant de
connecter les fils.
1. Use a Phillips-head screwdriver to remove the screws from the enclosure lid.
2. Remove the lid.
3. Run the AC power conduit/cord up through the bottom left opening on the
bottom of the enclosure.
4. Connect the conduit wires as described in Table 1and Figure 5.
5. Replace the cover.
6. Replace the cover screws.
FIGURE 5. AC wiring connections
Table 1. Aqua Con TROLL AC wiring connections
Pin Description North American Wire Color European Wire Color
LLine Black Brown
N Neutral White Blue
Protective Earth Shield Green and yellow
Ground
Neutral
Line
Pre-wired at factory
Run conduit through enclosure to terminals