Use Instructions
44
If the Vesta Charger and the OPTIMIZER Smart Mini IPG do not communicate
within a programmable time period, the patient may see the “Long Time Without
Downloading Data From IPG” alert screen displayed by the Vesta Charger:
In this case, instruct the patient to attempt to charge their OPTIMIZER Smart Mini
IPG with their Vesta Charger. If the patient is able to charge their implanted
device successfully, then the alert screen should no longer be displayed by the
Vesta Charger. If the attempt to charge the OPTIMIZER Smart Mini IPG with the
Vesta Charger is unsuccessful, the Impulse Dynamics representative should be
contacted.
Troubleshooting for Wireless Coexistence Issues
Troubleshooting OPTIlink Connection between the OPTIMIZER Smart Mini
IPG and the Intelio Programmer
If you experience issues with establishing an OPTIlink session between the
OPTIMIZER Smart Mini IPG and the Intelio Programmer, try the following:
• Reposition the Intelio Programming Wand so that it lays parallel to the
IPG’s plane and its center is coaxial with the center of the IPG’s header.
• Decrease the distance between the devices.
• Move the devices away from other devices that may be causing
interference.
• Do not operate other wireless devices (i.e., programmers for other
devices, laptop, tablet, mobile phone, or cordless phone) at the same
time.
If you experience issues with maintaining an OPTIlink session between the
OPTIMIZER Smart Mini IPG and the Intelio Programmer, try the following:
• Decrease the distance between the devices.
• Move the devices so they share line of sight.
• Move the devices away from other devices that may be causing
interference.
• Do not operate other wireless devices (i.e., programmers for other
devices, laptop, tablet, mobile phone, or cordless phone) at the same
time.
• Wait a few minutes and try connecting again
NOTE: Wireless communications equipment, such as wireless home network
devices, mobile and cordless telephones, and tablets, could affect the quality of
the OPTIlink connection.