Use Instructions

38
ISO 14117:2019
Clause 5 - Testing
above frequency of
3000 MHz
Standard does not
require testing of
devices above
3 GHz.
Electromagnetic
fields > 3 GHz are
not expected to
interfere with
device operation
because of the
increased device
protection
afforded by the
attenuation of the
enclosure and
body tissue at
microwave
frequencies, the
expected
performance of
EMI control
features
implemented to
meet lower-
frequency
requirements, and
the reduced
sensitivity of
circuits at
microwave
frequencies.
N/A
Avoid direct exposure to the main
lobe of high-power radar and
microwave communication beams.
ISO 14117:2019
Clause 6.1 - Protection
of the device from
damage caused by
high-frequency surgical
exposure
Per standard
Does not exhibit
malfunction which
persists after the
removal of the
electromagnetic
test signal per
standard
See section on Warnings
Electrocautery and Warnings
RF Ablation in this manual
ISO 14117:2019
Clause 6.2 Protection
of the device from
damage caused by
external defibrillators
Per standard
Does not exhibit
malfunction which
persists after the
removal of the
electromagnetic
test signal per
standard
See section on Warnings
Defibrillation and Cardioversion
in this manual