Use Instructions
37
ISO 14117:2019
Clause 4.8 - Protection
from AC magnetic field
exposure in the range
of 1 kHz to 140 kHz
Per standard
Does not exhibit
malfunction which
persists after the
removal from the
field per standard
See section on Cautions →
Environmental Conditions,
Cautions → Industrial
Machinery, and Cautions →
Home Appliances in this manual.
• Exercise caution in the
vicinity of equipment that
generates strong AC
magnetic fields.
• Do not enter an area with
posted warnings advising
pacemaker patients (or
patients with other types
of implantable devices)
not to approach.
ISO 14117:2019
Clause 4.9 - Test
requirements for the
frequency range of 385
MHz ≤ ƒ ≤ 3000 MHz
Per standard
Functions as it
did before the
test without
further
adjustment after
application of the
test signal per
standard
See section on Cautions →
Transmitting Devices and
Cautions → Cellular and Mobile
Phones in this manual
• Exercise caution in the
vicinity of equipment that
generates strong radio-
frequency fields.
• Do not enter an area with
posted warnings advising
pacemaker patients (or
patients with other types
of implantable devices)
not to approach.
• Interference may occur in
the vicinity of equipment
marked with the following
symbol: