Use Instructions

37
ISO 14117:2019
Clause 4.8 - Protection
from AC magnetic field
exposure in the range
of 1 kHz to 140 kHz
Per standard
Does not exhibit
malfunction which
persists after the
removal from the
field per standard
See section on Cautions
Environmental Conditions,
Cautions Industrial
Machinery, and Cautions
Home Appliances in this manual.
Exercise caution in the
vicinity of equipment that
generates strong AC
magnetic fields.
Do not enter an area with
posted warnings advising
pacemaker patients (or
patients with other types
of implantable devices)
not to approach.
ISO 14117:2019
Clause 4.9 - Test
requirements for the
frequency range of 385
MHz ≤ ƒ ≤ 3000 MHz
Per standard
Functions as it
did before the
test without
further
adjustment after
application of the
test signal per
standard
See section on Cautions
Transmitting Devices and
Cautions Cellular and Mobile
Phones in this manual
Exercise caution in the
vicinity of equipment that
generates strong radio-
frequency fields.
Do not enter an area with
posted warnings advising
pacemaker patients (or
patients with other types
of implantable devices)
not to approach.
Interference may occur in
the vicinity of equipment
marked with the following
symbol: