Use Instructions

34
V
BAT
(V)
Parallel Lead (V1+V2)
Impedance (Ω)
CCM Amplitude (V)
Average Measured
Current Drain (mA)
3.5
300
6
1.22
3.5
300
7.5
2.2
3.5
600
4.5
0.41
3.5
600
6
0.78
3.5
600
7.5
1.5
3.5
900
4.5
0.34
3.5
900
6
0.6
3.5
900
7.5
1.3
3.5
1200
4.5
0.31
3.5
1200
6
0.5
3.5
1200
7.5
1.2
4.1
220
4.5
1.21
4.1
220
6
1.46
4.1
220
7.5
2.13
4.1
250
4.5
0.7
4.1
250
6
1.42
4.1
250
7.5
1.8
4.1
300
4.5
0.68
4.1
300
6
1.08
4.1
300
7.5
1.47
4.1
600
4.5
0.52
4.1
600
6
0.65
4.1
600
7.5
1.06
4.1
900
4.5
0.38
4.1
900
6
0.46
4.1
900
7.5
0.97
4.1
1200
4.5
0.32
4.1
1200
6
0.48
4.1
1200
7.5
0.91