Use Instructions

33
Table 2
Parallel Lead (V1+V2)
Impedance (Ω)
CCM Amplitude (V)
Charge Longevity
(days)
220
4.5
38
220
6
25
220
7.5
16
250
4.5
52
250
6
29
250
7.5
19
300
4.5
57
300
6
36
300
7.5
22
600
4.5
88
600
6
57
600
7.5
32
900
4.5
113
900
6
77
900
7.5
36
1200
4.5
130
1200
6
83
1200
7
39
Battery Current Drain
The battery current drain of the OPTIMIZER Smart Mini IPG is highly dependent on the amount of
energy used when CCM therapy delivered to the patient.
Table 3 shows the average measured current drain from the OPTIMIZER Smart Mini IPG battery
during CCM therapy delivery under the following conditions:
Number of pulses per CCM train: 2
Phase duration: 5.13 ms
Heart rate: 75 bpm
100% CCM therapy delivery
Table 3
V
BAT
(V)
Parallel Lead (V1+V2)
Impedance (Ω)
CCM Amplitude (V)
Average Measured
Current Drain (mA)
3.5
220
4.5
0.96
3.5
220
6
1.84
3.5
220
7.5
2.9
3.5
250
4.5
0.88
3.5
250
6
1.36
3.5
250
7.5
2.4
3.5
300
4.5
0.75