Use Instructions
31
TIMING ALGORITHM
LS Assignment
V2
LS Alert Start
-10 ms
LS Alert Width
30 ms
LS BLANKING REFRACTORIES
Pre A LS Refractory Period
0 ms
Post A LS Refractory Period
0 ms
Pre RV LS Refractory Period
0 ms
Post RV LS Refractory Period
0 ms
Post LS Refractory Period
20 ms
CCM TRAIN
CCM Train Delay
30 ms
CCM Amplitude
7.5 V
Number of Biphasic Pulses
2
Balancing
40 ms
First Phase Polarity
Positive
Phase Duration
5.13 ms
Interval
0 ms
CCM Channels
RV, LS
PATIENT ALERTS (See Intelio Programmer System and Vesta Charger System IFU for more information)
Alert Delivery Mode
Scheduled 08 to 21
Max Lead Impedance Change
ON 30%
Minimum Target CCM Therapy Rate
ON 75%
Battery Recharge Reminder
ON
Battery Recharge Reminder days
10 days
CCM Therapy Suspended
OFF
Long Time Without Communicating With the IPG
ON 2 days
Long Time Without Transmitting Data to the Remote Monitor
OFF
Down Mode
ON
CCM Not Sensing/Noise
ON
Charger Battery Low
ON
Charger Failure
ON
Rechargeable Battery Low
ON