Use Instructions

30
CCM TRAIN PARAMETERS
Balancing
40 ms to 100 ms in 10 ms increments
First Phase Polarity
Positive” or “Negative.
Phase Duration
4 possible between 5.13 ms and 6.60 ms.
Interval
0 ms to 7 ms in 1 ms increments
CCM Channels
RV and/or LS
Default Settings
CCM THERAPY
Mode
OFF
Timed
1
7 hs/day
CCM Magnet Mode
Off 1 day
Extend on low CCM%
OFF
1
The default setting in the U.S. is 5 hs/day
CCM SCHEDULE
Start Time
00:00
End Time
23:59
SENSING
Atrium Sensitivity
1.3 mV
Atrium Polarity
Bipolar
Ventricle 1 Sensitivity
2 mV
Ventricle 1 Polarity
Bipolar
Ventricle 2 Sensitivity
2 mV
Ventricle 2 Polarity
Bipolar
A/V REFRACTORIES
Post-V Atrial Refractory Period
249.4 ms
Post-V Ventricular Refractory Period
249.4 ms
CCM INHIBIT
CCM Inhibit Cycles
2 beats
Short AV Delay
70 ms
Long AV Delay
397 ms
Tachycardia
1
98 bpm
1
Tachycardia controls the Atrial rate in ODO-LS-CCM mode and the Ventricular rate in OVO-LS-CCM mode