Use Instructions

29
A/V REFRACTORY PARAMETERS
Parameter Name
Values
Post-V Atrial Refractory Period
1
148.0 ms to 452.2 ms in 7.8 ms increments
Post-V RV Refractory Period
148.0 ms to 452.2 ms in 7.8 ms increments
1
Active only when the OPTIMIZER Smart Mini IPG is in ODO-LS-CCM mode.
CCM INHIBIT PARAMETERS
Parameter Name
Values
CCM Inhibit Cycles
1 to 16 in increments of 1
Short AV Limit
1
49 possible between 23 ms and 397 ms
Long AV Limit
1
49 possible between 23 ms and 397 ms
Atrial Tachycardia Rate
1
51 possible between 62 bpm and 179 bpm
Ventricular Tachycardia Rate
2
25 possible between 62 bpm and 110 bpm
1
Active only when the OPTIMIZER Smart Mini IPG is in ODO-LS-CCM mode.
2
Active only when the OPTIMIZER Smart Mini IPG is in OVO-LS-CCM mode.
CCM TIMING PARAMETERS
Parameter Name
Values
LS Assignment
V1 or V2
LS Alert Start
100 ms to 100 ms in 2 ms increments
LS Alert Width
1 ms to 40 ms in 1 ms increments
LS BLANKING REFRACTORY PARAMETERS
Parameter Name
Values
Pre A LS Refractory Period
1
0 ms to 55 ms in 5 ms increments
Post A LS Refractory Period
1
0 ms to 55 ms in 5 ms increments
Pre RV LS Refractory Period
0 ms to 55 ms in 5 ms increments
Post RV LS Refractory Period
0 ms to 39 ms in 1 ms increments
Post LS Refractory Period
15 ms to 250 ms in 5 ms increments
1
Active only when the OPTIMIZER Smart Mini IPG is in ODO-LS-CCM mode.
CCM TRAIN PARAMETERS
Parameters Name
Values
CCM Train Delay
3 ms to 140 ms in 1 ms increments
CCM Amplitude
4.5 V to 7.5 V in 0.5 V increments
Number of Biphasic Pulses
1, 2, or 3