Use Instructions
28
Safe Mode
Mode
Description
DOWN Mode
Occurs when the device encounters conditions considered to be the
result of faulty device hardware or firmware. In this mode, the device is
completely quiescent; CCM therapy is not delivered and cardiac
events are not sensed.
Programmable Parameters
DEVICE MODES
Mode
Characteristics
OOO
Standby mode: no events are sensed and no CCM pulse trains are
delivered
ODO-LS-CCM
Active mode where the device senses atrial, ventricular and Local
Sense events and is capable of CCM therapy delivery
OVO-LS-CCM
Active mode where the device senses ventricular and local sense
events and is capable of CCM therapy delivery without the need for
the detection of atrial sense events.
CCM THERAPY PARAMETERS
Parameters Name
Values
CCM Therapy Mode
OFF
No pulse train enabled
ON
As defined by the parameter values below
CCM Therapy (hs/day)
1 hs/day to 24 hs/day in 1 hs/day increments
Start Time (hour)
00 h to 23 h in 1 h increments
Start Time (minute)
00 m to 59 m in 1 m increments
End Time (hour)
00 h to 23 h in 1 h increments
End Time (minute)
00 m to 59 m in 1 m increments
CCM Magnet Mode
Off 1 day or Off
Extend on Low CCM%
On or Off
A /V SENSING PARAMETERS
Parameter Name
Values
Atrium Sensitivity
1
11 possible between 0.3 mV to 5 mV
Atrium Polarity
1
Bipolar or Unipolar
Ventricle 1 Sensitivity
16 possible between 0.3 mV and 10 mV
Ventricle 1 Polarity
Bipolar or Unipolar
Ventricle 2 Sensitivity
16 possible between 0.3 mV and 10 mV
Ventricle 2 Polarity
Bipolar or Unipolar
1
Active only when the OPTIMIZER Smart Mini IPG is in ODO-LS-CCM mode.