Use Instructions
26
9. Post LS Refractory period should not be greater than the CCM Train Delay
Rationale: If the CCM Train Delay is shorter than the Post LS Refractory period, then the
CCM therapy will be delivered within the Post LS Refractory period while the LS event is
not sensed.
11.0 SERVICE AND WARRANTY
11.1 Limited Warranty Information
Impulse Dynamics warrants that all IPGs (including the respective firmware and software) will be
free from defects in workmanship and materials for a period of 24 months after the original
implantation of the IPG, unless a longer period is required pursuant to applicable law (the
“Warranty Period”).
If it appears that any IPG or part thereof appears to be defective in workmanship or materials, or
fails to conform to applicable specifications, Impulse Dynamics shall either replace defective or
non-conforming implantable components or repair or replace defective or non-conforming non-
implantable components. The warranty period for a replaced or repaired IPG shall be the time
remaining on the original warranty period or nine months from delivery of the repaired or replaced
IPG, whichever is longer.
Under this warranty, Impulse Dynamics shall not be liable if tests and analyses reveal that the
alleged defect or non-conformity of the IPG is not present or was caused by improper use,
neglect, improper implantation, or follow-up, unauthorized repair attempts by the user, or due to
accident, fire, lightning, or other hazards.
11.2 Mandatory Battery Charging
The rechargeable battery in the OPTIMIZER Smart Mini IPG is designed to provide optimal
performance if it is completely recharged on a weekly basis. Regular weekly recharging sessions
are required to prevent battery deterioration, which may lead to decreased device longevity.