Use Instructions

9
5.6 Therapeutic Ultrasound
Warning: Direct exposure of the OPTIMIZER Smart Mini IPG to therapeutic ultrasound can
damage the device. In addition, the OPTIMIZER Smart Mini IPG can inadvertently
concentrate the ultrasound field and cause harm to the patient.
Therapeutic ultrasound can be used provided the implant is located far away from the ultrasound
field. Precautionary programming of the OPTIMIZER Smart Mini IPG to OOO mode reduces the
risk of adverse effects. The patient’s peripheral pulse should be monitored during the procedure.
Immediately after the treatment, the OPTIMIZER Smart Mini IPG should be checked for proper
function. If the device is found to have reverted to its “DOWN” mode, it needs to be reset.
5.7 Nuclear Magnetic Resonance (NMR), Magnetic Resonance Imaging
(MRI)
Warning: Patients with an implanted OPTIMIZER Smart Mini IPG should not be exposed to NMR
or MRI.
Exposure of the OPTIMIZER Smart Mini system to strong magnetic and electromagnetic fields
encountered within MRI systems has not been investigated. Even though programming the IPG
into Standby (OOO) mode reduces the risk of adverse events, exposure of the patient to an MRI
scan could result in the following:
Unintended cardiac stimulation (induced tachycardia)
Tissue damage near the IPG and lead electrodes with the result of inability to deliver
CCM therapy
Device malfunction (discharge of battery, damage to device electronics, reverted to its
“DOWN” mode)
If the device is found to have reverted to its “DOWN” mode, it will need to be reset, which will
clear the statistical data stored in the device.
5.8 Radiation Therapy
Warning: Therapeutic equipment generating ionizing radiation, such as linear accelerators and
cobalt machines employed for treating malignant diseases, can damage the circuits
used in most active implantable devices. Because the effect is cumulative, both dose
rate and total dose determine if damage will occur and its possible extent. Be aware
that certain types of damage might not be immediately detectable. In addition, the
electromagnetic fields generated by some types of radiation equipment for beam
“steering” purposes can affect the function of the OPTIMIZER Smart Mini IPG.
Radiation therapy can lead to a wide spectrum of effects, reaching from transient interference to
permanent damage. It is therefore advisable to locally shield the OPTIMIZER Smart Mini IPG
against radiation if radiation therapy is to be used. During a radiation treatment and thereafter, the
function of the IPG needs to be monitored. If tissue in the vicinity of the implant has to be
irradiated, it may be advisable to relocate the OPTIMIZER Smart Mini IPG.
5.9 Lithotripsy
Warning: Direct exposure of the OPTIMIZER Smart Mini IPG to shock waves can damage the
device. A device implanted outside the shock wave path presents no clear-cut
contraindication to lithotripsy. Precautionary programming of the OPTIMIZER Smart
Mini IPG to OOO mode reduces the risk of adverse effects. The patient’s peripheral
pulse should be monitored during the procedure. Immediately after the treatment, the
OPTIMIZER Smart Mini IPG should to be checked for proper function. If the device is
found to have reverted to its “DOWN” mode, it will need to be reset by qualified
personnel. A consequence of a device reset is that all statistical data stored in the IPG
is cleared.