Use Instructions

7
5.1.3 Atrial Arrhythmias Potentially Caused by CCM Therapy Pulses
Atrial and supraventricular arrhythmias could theoretically be initiated when CCM-induced
ventricular activity is conducted retrograde to the atria, resulting in premature atrial
depolarization. The OPTIMIZER Smart Mini IPG may sense the ventricular activation
resulting from the retrograde-induced atrial event and deliver CCM therapy as
programmed. In addition, strong CCM therapy pulses delivered through leads implanted
in a basal position close to the atria have the potential of directly stimulating the atria. If
CCM therapy causes atrial activation through either of these mechanisms, and the atrial
signal is then conducted to the ventricles, it may look like couplet PACs (AVAV) but the
second complex would be identified as a “PVC” by the OPTIMIZER Smart Mini IPG.
The main variables that may cause CCM therapy pulses to lead to atrial activation are
ventricular lead placement location on the right ventricular septum, CCM pulse amplitude,
and CCM train delay. To prevent the occurrence of atrial arrhythmias due to CCM
therapy pulses, it is recommended that basal lead implant locations be avoided.
The potential for direct atrial activation by CCM therapy pulses can be tested during
implant by setting the CCM pulse amplitude to the highest possible value and extending
the CCM train delay by 20 to 30 ms beyond its recommended setting, while ensuring that
the CCM therapy pulse train, including its balancing phase, remain completely within the
bounds of the ventricular absolute refractory period, then delivering CCM therapy while
monitoring the patient’s heart rhythm for episodes of atrial activation. The testing should
confirm the absence of atrial activation with the increased CCM amplitude and extended
CCM train delay.
In addition to proper lead location and CCM parameter programming, another protective
measure that must be implemented is the programming of the Atrial Tachycardia Rate
(ODO-LS-CCM mode only) to a sufficiently low value to prevent CCM therapy delivery
from inducing atrial arrhythmias while still allowing for the consistent delivery of CCM
therapy.
5.2 Electrocautery
Warning: The use of surgical electrocautery devices, especially of the unipolar kind, can induce
CCM therapy inhibition or can make the OPTIMIZER Smart Mini IPG revert to its
“DOWN” mode (OOO mode, with no delivery of CCM). If the device is found to have
reverted to its “DOWN” mode, it will need to be reset, which will clear the statistical
data stored in the device. The device can be damaged if high energies are coupled
into the system.
Use of electrocautery in close proximity to an implanted OPTIMIZER Smart Mini IPG can also
couple radiofrequency (RF) energy directly through the leads and lead tips into the cardiac
muscle tissue, producing burns or possibly cardiac arrhythmias. If electrocautery is used, brief
signal bursts should be considered, with the neutral electrode positioned such that its affects on
the OPTIMIZER Smart Mini IPG and its attached leads are minimized. The risk of adverse effects
can be decreased by reprogramming the OPTIMIZER Smart Mini IPG into OOO mode. The
patient’s peripheral pulse should be monitored throughout the procedure and correct operation of
the OPTIMIZER Smart Mini IPG should be verified immediately after the procedure.