Use Instructions

6
Lead fracture
Lead displacement
Atrial or ventricular perforation
Rare cases of pericardial tamponade
Perforation of the ventricular wall can induce direct stimulation of the phrenic nerve or the
diaphragm. An impedance change demonstrated on a check-up can be indicative of lead fracture,
lead displacement, lead insulation damage, or perforation (also see: Potential Adverse Effects,
Section 7).
In very rare cases (<1%), transvenous lead placement can lead to venous thrombosis and
subsequent SVC syndrome.
Loss of sensing shortly after implant can be the result of lead displacement. In addition, loss of
CCM therapy delivery could be due to a lead fracture.
5.1.1 Atrial and Ventricular Arrhythmias Potentially Caused by Lead Implantation
The use of transvenous leads may lead to arrhythmias, some of which may be life-
threatening, such as ventricular fibrillation and ventricular tachycardia. The use of screw-
in leads, such as those used for CCM therapy delivery, has the potential of causing
conduction disturbances such as bundle branch block. These can be minimized by
performing the implant under fluoroscopic guidance, ensuring that the leads are placed in
the appropriate position prior to fixation, as well as limiting the number of lead
manipulations.
Please read and follow all directions in the Instruction for Use document provided with the
leads that you intend to use in order to minimize adverse events associated with lead
implantation.
5.1.2 Ventricular Arrhythmias Potentially Caused by CCM therapy pulses
CCM therapy pulses are of greater strength than that of typical pacing pulses and are
thus capable of eliciting activation of cardiac tissue when delivered outside of the
absolute refractory period. CCM therapy pulses delivered outside of the ventricular
absolute refractory period thus have the potential of causing pulse-induced arrhythmias
(some of which may be life-threatening, such as ventricular fibrillation and tachycardia).
For this reason, it is imperative that CCM therapy delivery parameters be chosen
carefully. Most importantly, the various settings related to conditions that inhibit CCM
therapy delivery (e.g., Long AV Delay, Short AV Delay, LS Alert Window, refractory
periods, and IEGM sensitivities) must be selected to allow delivery of CCM therapy only
on normally conducted (e.g., non-arrhythmic) beats, but inhibit them on beats of
suspected ectopic or premature origin.
In addition, CCM therapy pulses may cause changes in the electrical conduction of
tissue. For this reason, the delivery of CCM therapy pulses to the ventricular septum has
the potential of causing bundle branch block that could lead to bradycardia. Through
similar mechanisms, CCM-induced changes in the electrical conduction of the
myocardium have the potential of inducing tissue refractoriness that may facilitate the
induction of reentrant tachyarrhythmias. It is recommended that the patient be monitored
carefully for changes in heart rhythm when CCM therapy is temporarily activated during
lead implantation, as well as during the first permanent activation of CCM therapy after
implant and subsequent follow-up visits. Changes in ventricular rhythm caused by the
delivery of CCM therapy pulses may require repositioning the leads, and/or changing the
CCM train delay and CCM amplitude parameters to values that do not adversely affect
the patient’s ventricular rhythm.