Use Instructions

2
Pacing Leads Suitable for use with the OPTIMIZER IPG for CCM Signal Delivery
Current Offerings as of January 2, 2019
Requirement
for CCM
Medtronic
CapSureFix
Novus MRI™
SureScan™
4076, 5076,
5086 Leads
Abbott
(St Jude)
2088TC
Tendril
STS lead
Abbott
(St Jude)
LPA1200M
Tendril MRI
Lead
Boston
Scientific
Ingevity
7740, 7741,
7742
Leads
Biotronik
Solia-S
Leads
Bipolar lead
approved for
transvenous
intracardiac
ventricular
pacing
YES
YES
YES
YES
YES
Standard IS-1
bipolar
connector
YES
YES
YES
YES
YES
Active fixation
with electrically-
active
corkscrew distal
electrode with a
minimal
electrically-
active surface
area of 3.6 mm
2
YES,
4.2 mm
2
YES,
6.9 mm
2
YES,
6.0 mm
2
YES,
4.5 mm
2
YES,
4.5 mm
2
Distal electrode
coated with low-
polarization
coating (e.g.
titanium nitride
or iridium oxide)
YES,
titanium nitride
coating
YES,
titanium
nitride
coating
YES,
titanium
nitride
coating
YES,
IROX™
(iridium
oxide)
coating
YES,
“Fractal
Iridium”
(iridium
oxide)
coating
1.3 OPTIMIZER Smart Mini IPG Lead Connectors
The connector block accepts three (3) bipolar IS-1-BI connectors. The terminals are marked as
follows:
“A”: Atrium In 2-Lead mode operation, the “A” port is plugged with a silicone plug
provided with the IPG.
“V1”: Ventricle 1
“V2”: Ventricle 2