Guidance and manufacturers declaration – electromagnetic immunity
The XD2000 PX is intended for use in the electromagnetic environment specified below.
The customer or the user of the XD2000 PX should assure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5 cycles
70 %UT
(30 %dip in UT)
for 25 cycles
<5%UT
(>95%dip in UT)
for 5 s
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5,6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the ISW-
MKR101 requires continued
operation during power mains
interruptions,
it is recommended that the
ISW-MKR101 be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.