User Manual
7
HS_KS 501 digital_092012
The attachments are placed on the
shaking table and fastened with the
lateral quick-action vises.
Separating funnel attachments are only
suitable for HS 501 digital.
Maintenance and Cleaning
The instrument requires no maintenance. Only the ball bearings are subjected to a
natural wear. If the instrument needs to be repaired, it may only be opened by a
qualified technician. If the instrument does not seem to function, please check the
fuses first: Disconnect mains plug of instrument and open fuse holder. For safety data
see Technical Data/Mains Connection.
The surfaces of the instrument may be cleaned with:
a) water with a tenside containing detergent
b) isopropyl alcohol (e.g. if very soiled) with the exception of the polystyrole attachments.
Please submit for repair only instruments that have been cleaned and are free from
substances dangerous to health.
When ordering spare parts, please indicate type of instrument, serial number
and ordering number of the spare part.
Technical Data
Mains voltage 230 V or 115 V
mains frequency 50/60Hz
mains connection GEE-instrument plug with fuses:
230 V = 2 x 0,5 A; 115 V = 2 x 1 A
protection class (DIN 40 050) IP 21
permissible duration of operation 100%
timer 0-56 min/co
power input 70 watts
power output 19 watts
speed 0-300 1/min
speed display digital
shaking motion HS 501 to and fro
KS 501 orbital
max. load incl. attachment 15 kg
permissible ambient temperature +5°C - +50°C
relative humidity max. 80%
dimensions W x D x H 505 x 585 x 120 mm
weight (without attachment) 24 kg
Subject to technical changes!