Manual

19
Error codes
An error is indicated by an acoustic signal and an error code is displayed.
E. g.:
If faults cannot be eliminated directly, you must perform a RESET (see „Resetting parameters to the factory settings“)!
If the faults still cannot be eliminated, the appliance must be inspected by a technical service.
Error Description Cause Effect Solution
code
Er 3 Temperature inside unit too high Permitted ambient temperature has Heating off • Switch off the unit. Allow it to cool down
been exeeded and switch on again.
Ventilation slots or fan housing • Clean ventilation slots or fan
blocked housing
• Observe maximum permissible ambient-
temperature
Er 4 Differance between setpoint and Motor seized or overload Motor off Reduce the torque load
actual speeds is too large (load)
• Reduce setpoint speed
Er 8 The calibration value of the tempe- Fault in calibration procedure Heating off • Repeat the calibration procedure
rature sensor is outside the limit Value was incorrectly stored to memory
value EPROM switching error
Er 14 External temperature sensor, Short-circuit at temperature sensor Heating off • Check the plug
short-circuit plug • Substitude the temperature sensor
Short-circuit in connecting cable
or on Temperature sensor
Er 16 External temperature sensor has SAFE Temp is set at a temperature Heating off Allow the unit to cool down
exceeded the SAFE Temp lower than the actual temperature on • Set the SAFE Temp at a higher temperature
the external temperature sensor
Er 17 Temperature sensor - Incubation- SAFE Temp is set at a temperature Heating off Allow the unit to cool down
room has exceeded the lower than the actual temperature on • Set the SAFE Temp at a higher temperature
SAFE Temp the external temperature sensor
Er 26 Difference between the internal Ventilation slots in incubation room Heating off • Switch off the unit. Allow it to cool down
control and safety temperature blocked and then switch on again.
sensors too large Radial-ow fan does not rotate • Check fan or ventilation slots and
Fault in the control or safety clean if necessary
temperature sensors
Er 60 Power outage Power outage during operation Interruption of heating • Delete the display by pressing the
or shaking function shift key
PC 1 In remote operation (PC) with •PC does not send data during the Heating off • Change the watchdog time
watchdog function 1 (WD) enabled: watchdog time Motor off Send data from the PC within the
No communication between The connection/cable to the PC is watchdog time
PC and unit broken Check the connector cable and plug
PC 2 In remote operation (PC) with PC does not send data during the The setpoint temperature • Change the watchdog time
watchdog function 2 (WD) enabled: watchdog time is set to the WD safety • Send data from the PC within the
No communication between temperature watchdog time
PC and unit The connection/cable to the PC is The setpoint speed • Check the connector cable and plug
broken is set to the WD safety
speed