Owner manual
Error code
Error Cause of the error
Off - LED (red) The potentiometer for -Too rapid a change in the
1x flash / the speed is not assig- speed target value
2 sec pause ning any target value -Internal device error
Off - LED (red) Permissible motor tem- -Permissible supported
2x flash / peratur exceeded weight exceeded
2 sec pause -Permissible ambient
temperature exceeded
Motor locked -Agitation table is being
hindered in this stroke mo-
tion by an outside effect
Read fork signal not -Internal device error
available
Off - LED (red) Maximum adjustable -Resonance behavior of
3x flash / speed of the device the experiment setup
2 sec pause exceeded
Off - LED (red) Motor standstill for de- -When turning on the
4x flash / vice switch-on proce- device, you must ensure
2 sec pause dure (safety confirma- that the motor is not
tion not ensured performing any rotational
motion
Off - LED (red) Safety relay cannot -Internal device error
5x flash / be activated
2 sec pause
Off - LED (red) The switch for the TIME -Internal device error
6x flash / function is not assigning
2 sec pause any target value
Off - LED (red) Problem with the power -The device is beeing
7x flash / supply operating with under vol-
2 sec pause tage or overvoltage (for
permissible voltage range,
see technical data)
Off - LED (red) Triac cannot be activated -Internal device error
8x flash /
2 sec pause
29
032007
place evenly on the agitation table, and that all vessels are secu-
rely fastened in place.
Basic error codes:
ATTENTION! If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the dis-
play field. First, please try to continue operation by turning the
device off and back on again.If it is still not possible to eliminate
an error after an extended wait, please contact our service
department. If you do so, you should always tell us what error
code has been encountered. This simplifies the process of
tracking down the error and makes it possible to form a prelimi-
nary opinion.
AS 130.1 Universal attachment
AS 130.4 Reactant glass adapter
1/min
0 80 560
160
240 320 400 480 640 800720
g
2000
1750
1500
1000
500
0
1250
750
250
1/min
0 80 560
160
240 320 400 480 640 800720
g
2500
2000
1500
1000
500
0
Error codes