User Manual

12
CMAG dig 0511
Commis-
sioning
Put device switch (A) in the OFF position
Plug in (G) mains cable
Once connected to the power supply the device
is in “stand-by” mode
Put device switch (A) in the ON position
Set the engine speed with the operating button (MOT)
Any set values are retained when device is switched
off and even after the device is disconnected from the
power!
Commissioning
Stirring
Heating
without external
temperature sensor
not in
operating mode A
with external
temperature sensor
(direct
temperature
control in the
medium)
Put device switch (A) in the OFF position
Plug in the PT1000 sensor into the socket (I)
Put device switch (A) in the ON position
Adjust the set-point temerature
with the operating but-
ton
(TEMP)
LED (F) lights up
the instantaneous temperature (K) is indicated on the
display (corresponds sensor- and/or medium temperature)
indicated set point temperature (L) corresponds the
desired medium temperature
LED (E) lights up and the symbol for energy input (M) is
indicated on the display (with the heating is switched on)
indicate HOT (D) on the display with over 50°C on the
heating plate (in the agitating and stand-by-mode)
Put device switch (A) in the ON position
Set the operating mode (see capture “Operating
modes”)
Adjust the set-point temperature with the operating but-
ton (TEMP) (only be set in operating mode B)
indicate the adjusted set point temperature on the
LCD display (D) position (L)
indicated the set point temperature corresponds the
heating plate temperature
LED (E) lights up and the symbol for energy input
(M) is indicated on the display (with the heating is
switched on)
indicate HOT (D) on the display with over 50°C on
the heating plate (in the agitating and stand-by-
mode)
Any malfunctions during operation will be identified by an error mes-
sage on the display (D).
Proceed as follows in such cases:
Switch off device using the main switch (A)
Carry out corrective measures
Restart device
Error codes
Error Cause Solution
code
E1 Sensor not • Connect sensor
connected • Change operating mode
(operating mode A)
E3 Temperature inside Switch off device and allow
device too high to cool down
E5 Sensor not in • Sensor in medium
media immerse
E6 Sensor defectively • Exchange the sensor
E7 Hi-Temp • Medium to cool down
temperature border
exceeded
E9
Initializations error
• Switch off/on the device
E11 Motor blocks use other stirring bar
• reduce viscosity
If the actions described fail to resolve the fault or another error
code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the
fault.