It is only possible to calculate net calorific values and publish them by printing or on
the LCD-display after all pre-test and post-test parameters have been entered DQG
the measurements carried out. The output contains all the parameters entered, the
automatically recorded data, and the calculated results. The latter are dependent on
the calculation mode selected. In addition, the output contains notes about the test,
if appropriate.
To print test results, proceed as follows:
c
Make sure that the printer is connected and switched on.
d
Change from PHQX to the sub-menu RXWSXW by pressing the function button I, and
then use function button I to select the sub-menu VDPS
e
Enter selection criteria for the date, sample code and sample status as described
above, and close your entries by pressing the PHP button. The printer will now start
to print.
To show test results on the LCD-display, proceed as follows:
,IWKHUHLVQRSULQWHUFRQQHFWHGRULI WKHSULQWHULVQRWVZLWFKHGRQ\RXFDQ
RQO\VKRZWHVWGDWDRQWKH/&'GLVSOD\LIWKHSDUDPHWHUUHDOWLQWKHVXEPHQX
67,QLWRIPHQXLVVHWWR
c
Change from PHQX to the sub-menu RXWSXW by pressing the function button I, and
then use function button I to select sub-menu /&'.
d
Enter selection criteria for the date, sample code and sample status as described
above, and close your entries by pressing the PHP button. The parameters appear
in the LCD-display in two lines, in the same order as the printed output. The upper
line contains the name and units of the parameter, and its value is shown in the
lower line. To see the next parameter, use the arrow buttons
or
(in this case,
both have the same effect). After all parameters have been shown, the display re-
turns to the main menu.
Parameters can only be displayed in the forwards direction, it is not possible to page
backwards. The HVF button can be used at any time to break off the procedure.
(GLWLQJWHVWSDUDPHWHUV
The subsequent editing (changing) of parameters is only possible if the test results
have been issued at least once by printing or on the LCD-display. This is to prevent
values being changed unintentionally.