Instruction Manual

IKA
-
WERKE C 7000 Ver. 07 09.07
3DJH
The cooler is used to cool the decomposition vessel quickly after a test. For this
p
urpose, the cooler C 7002 is connected using the cable supplied to the calorimeter
C 7000 (see illustrations Cooler C 7002 Rear view, Position 2, and C 7000 Rear
view, Position 3). During the cooling operation, the decomposition vessel is cooled
to the temperature of the measuring cell. For accurate calibration and measurement,
this condition must be fulfilled.
For cooling, the decomposition vessel is placed between the cooling jaws. Peltier
elements are used for cooling. They are controlled completely automatically. Water
is employed for removal of heat. Internal controls ensure the water consumption is
kept especially low. In stand-by mode, the water is switched off automatically. The
cooler does not start ist pre-cooling phase until a main test has started.
Please see Section 6.2 "Connection of Peripherals" for more information about in-
stallation of the cooler.
&RROHU
&
5HDUYLHZ
1 Mains connection
w
ith main fuse
2 Connection to
C 7000
3 Cooling water con-
nections with quick-
connect couplings