IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
+&DO
The experiment was performed for calibration purposes.
6LP
The experiment was a simulation.
(YDO
The experiment has been evaluated.
– Code Eval: the DIN/IKA evaluation procedure was used.
– Code ASTM: the ASTM D1989, D240, D5865, D4809, D5468,
E711 evaluation procedure was used.
:DLW
The fuel sample is in the crucible and all parameters have been en-
tered. The experiment can be started.
3UHS
The crucible is in the sample rack with a fuel sample.
5XQ
The experiment is currently being performed in the measurement cell.
The buttons have the following functions:
6HO Marks an experiment in the list
exception: calibration.
$OO Marks all experiments (max. 100 experiments) in the list.
Exception: Experiments with status FDO and 3UHS
3UL Prints the experiment list.
'HO Deletes experiments that have been previously selected with 6HO
or $OO. Exception: FDO3UHS
,QIR Opens an information window with the experiment parameters.
&DOF Opens a dialog box to convert the results of the experiment into vari-
ous reference states.
Using the 8SDUURZ'RZQDUURZ keys, you can select the experiment from the list
that you would like to post-process. Then you can move the cursor with 7DE from
the list box to the buttons.
You can activate the buttons either by placing the cursor on them with 7DE and
pressing 2. or by pressing the corresponding button number on the numeric key-
pad.
e
Place the cursor on the experiment that has just been completed and activate the
,QIR button. An information window appears with the results of the experiment.