Instruction Manual

28
8.4 Maintaining the water circulation
In the case of discontinuous use with tap water
(single measurements with long gaps) a stabiliser
must be added to the water circulation to prevent
the formation of algae.
Add approx. 4 ml IKA
®
-Aqua-Pro C 5003.1 to the
tank (see Accessories).
Switch the pump on via the service menu (sec-
tion “Pump“). Switch the pump off again after
30 - 60 s.
You can also add the 4 ml Aqua-Pro to the water
maintained at a constant temperature in the last
measurement (before long gaps in measuring).
If the appliance is not going to be in opera-
tion for a long period of time, it is advisable
to completely empty the calorimeter's water
circulation.
The water must be drained out before trans-
portation.
Empty the outer vessel via the service menu (sec-
tion 7.2 “Outer vessel“). The emptying hose must
be locked in place in the plug-in coupling (sec-
tion 4.2).
Empty the tank by locking the emptying hose in
place in the plug-in coupling (section 4.2). The
tank will empty automatically.
Press the locking knob on the plug-in coupling
to remove the hose.
8.5 Decomposition vessels
Please see the operating instructions for
C 5010 for information on decomposition
vessel maintenance!
8.6 Cleaning information
Only clean IKA
®
appliances using these IKA
®
ap-
proved cleaning agents:
Dirt Cleaning agent
Dyes Isopropanol
Building materials Water containing
detergent,
Isopropanol
Cosmetics Water containing
detergent,
Isopropanol
Food Water containing
detergent,
Fuels Water containing
detergent,
Other materials Please consult IKA
®
For cleaning the micro lter you must empty
the outer vessel (see section 7.2).
Screw the cover of the micro lter down and
pull you out the fi lter element.
Clean the fi lter element with clear water.
• Insert the lter element after cleaning with the
opening inward again into the housing of the
micro fi lter and screw you on by hand the co-
ver.
Fill the outer vessel again with water (see sec-
tion 4.7).
Control the tightness after you the equipment
again fi lled (see section 6.6 “Fill time exceeded).
TIP