User's Manual
14
/
14
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Table4–Inputa.c.powerPort
Phenomenon Basic EMC standard Immunit
y
test levels
Home Healthcare Environment
Electrical fast
transients/burst
IEC 61000-4-4 ±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6 3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands between 0.15MHz
and 80MHz
80%AM at 1kHz
Voltage dips IEC 61000-4-11 0% U
T
; 0.5 cycle
At 0º,45º,90º,135º,180º,225º,270º and 315º
0% U
T
; 1 cycle
and
70% U
T
; 25/30 cycles
Single phase: at 0º
Voltage interruptions IEC 61000-4-11 0% U
T;
250/300 cycles
Date of issue: Jul 4, 2017
BP5S-SMSY01 V1.0