ISO81060-2 2013,(Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of
Automated Measurement Type).
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This product is applicable to the equipment and system requirements for the purpose of
receiving radio frequency energy for the purpose of the work, Bluetooth receive bandwidth
2M. This product can also be used to include RF transmitter equipment and system
requirements and emission frequency of 2.4GHz ISM band, Bluetooth modulation
types:GFSK, effective max radiated power: <4dBm
Table 1 - Emission
Phenomenon Compliance Electromagnetic environment
RF emissions CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic distortion IEC 61000-3-2
Class A
Home healthcare environment
Vo l t a g e f l u c t ua t i o n s
and flicker
IEC 61000-3-3
Compliance
Home healthcare environment
Table 2 - Enclosure Port
Phenomenon Basic EMC standard Immunity test levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2 ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field IEC 61000-4-3 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Table 3 Proximity fields from RF wireless communications
equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780