TECHNICAL SPECIFICATIONS
Application
Measuring method
Oscillometric / non-invasive
Application
For adult use only
Measurement
location
Wrist section of lower arm
Memory
Maximum 30 records per user
Maximum number
of users
2
Dimensions
Height x Width x
Depth
3.5 x 3.4 x 1.2 inches
(90 x 86 x 30.8 mm)
Weight
6.3 oz (180g) without batteries
Cuff circumference
5.3 to 7.7 inches (13.5 to 19.5 cm)
Measuring range
Pressure
30 to 280 mmHg (4.0 to 37.3 kPa)
Pulse
40 to 200 pulse/min
Accuracy
Pressure
+/- 3 mmHg (+/- 0.4 kPa)
Pulse
+/- 5%
Error
message
EExx
appears
on the
display
A calibration
error
occurred
Retake the
measurement.
If the problem
persists, contact
the retailer or our
customer service
department
for further
assistance. Refer
to the warranty
for contact
information and
return instructions
Settings
are
wrong
Date and
time are
incorrect
The clock
was not set
or reset after
installing new
batteries
Reset the clock
No user
memory
Cannot
nd a
record
The memory
was not
saved after
taking a
measurement
Retake the
measurement.
Press
or
to save the
corresponding
user’s memory
The record
was stored
to the wrong
user memory
Press to view
User 1’s record.
Press
to view
User 2’s record
BPW800_EN_R8.indd 13 2011/6/8 7:07:11 PM