en es fr de Instructions for Use Instrucciones de uso Mode d’emploi Gebrauchsanleitung D R AF T Blood Pressure Monitor BP5000 BT
Contents 1. Introduction ...................................................................................................... 1 2. List of parts....................................................................................................... 2 3. Overview of symbols and functional features ................................................... 3 4. Setting up your blood pressure monitor ........................................................... 5 4.1 Installing the batteries...........................
1. Introduction The Blood Pressure Monitor BP5000 BT is a device that you will use to measure your blood pressure and report the results to your care provider. The blood pressure monitor BP5000 BT connects wirelessly to your Bosch telehealth appliance, such as the Health Buddy or the T400. This device is intended for use on adults aged 18-years and older. Read the entire instruction manual before you use your blood pressure monitor.
2. List of parts Ensure that your Bosch blood pressure monitor package includes the following items: Arm Cuff Bosch BP5000 BT Blood Pressure Monitor Four AA Batteries Keep the plastic packaging, the batteries, and the thin plastic film protecting the120 screen of the blood120 pressure monitor away from babies and children — 80 they80are a choking hazard. SYS mmHg EC 72 72 DIA mmHg PUL /min F.03D.601.424 Rev. A 12/2012 | Robert Bosch Healthcare Systems, Inc.
3. Overview of symbols and functional features Your BP5000 BT blood pressure monitor has a number of symbols and features. Memory Appears when your previous blood pressure measurements are displayed. Bluetooth Appears when the blood pressure monitor is connected to your telehealth appliance. Battery 3 | en Pulse Volume Start/Stop button Blinks when your pulse is being measured. Robert Bosch Healthcare Systems, Inc. | F.03D.601.424 Rev.
Volume button Bluetooth On / Off* Memory button Press to see and hear your last blood pressure measurement. Bluetooth pairing mode* Reset button* Language button By default, your blood pressure monitor is set to English. Battery ribbon Use to remove batteries *Do not press the Bluetooth or Reset buttons unless you are asked to do so by your care provider. For a complete explanation of the symbols and functional features of your blood pressure monitor, see Section 7. F.03D.601.424 Rev.
4. Setting up your blood pressure monitor Before you use your blood pressure monitor, you need to install the batteries and connect the arm cuff to the blood pressure monitor. Next, you need to set up the wireless connection between your blood pressure monitor and your Bosch telehealth appliance. Language button 4.1 Installing the batteries To install the batteries: 1. Open the battery compartment cover on the bottom of the blood pressure monitor. 2.
4. Replace the battery compartment cover. When you replace the batteries, ensure that you replace the complete set of batteries and that you do not use rechargeable batteries, or the battery fluid might leak. 4.2 Connecting the arm cuff to your blood pressure monitor To connect the arm cuff to your blood pressure monitor, plug the air hose into the air socket on the side of the blood pressure monitor. not use in combination with magnetic resonance imaging (MRI) . equipment.
4.3 Setting up the wireless connection To set up the wireless connection between your blood pressure monitor and your Bosch telehealth appliance: 1. Place your blood pressure monitor near your telehealth appliance to ensure a strong wireless connection. 2. If your care provider has provided you with a Bluetooth USB adapter, plug the Bluetooth USB adapter into any available USB port on the back of your telehealth appliance.
5. Measuring your blood pressure You can wait until your telehealth appliance prompts you to measure your blood pressure, or you can measure your blood pressure at any other time as directed by your care provider. Regardless of when you measure your blood pressure, the basic procedure is the same. Before you begin To ensure accurate blood pressure measurements, observe the following suggestions: • Try to measure your blood pressure at the same time each day.
• When the arm cuff inflates, it will become fairly tight around your arm. This is normal. Remain still until the measurement is complete. Procedure • The blood pressure cuff will automatically deflate if the cuff pressure exceeds 300mmHG. If the cuff pressure becomes too uncomfortable and does not automatically deflate, detach the cuff from your arm and contact your care provider. • If you have had a mastectomy or a dialysis graft, do not put the arm cuff around the affected side.
2. Fasten the blood pressure cuff around your upper arm about 1 inch above your elbow. 3. If you have a session waiting for you to complete on your telehealth appliance, start the session and follow the instructions on your telehealth appliance as you normally do. When you are prompted to measure your blood pressure, press the START button on your blood pressure monitor. The arm cuff inflates. F.03D.601.424 Rev. A 12/2012 | Robert Bosch Healthcare Systems, Inc.
When the measurement is complete, your blood pressure and pulse are displayed, and the arm cuff deflates automatically. If you do not have a session waiting for you to complete on your telehealth appliance, and you want to measure your blood pressure now, press the START button on your blood pressure monitor. The arm cuff inflates. When the measurement is complete, your blood pressure and pulse are displayed, and the arm cuff deflates automatically. 4. Remove the arm cuff from your arm.
6. Error messages on your blood pressure monitor When you use your blood pressure monitor, it might display an error message to indicate that a problem occurred. The following table provides explanations of the error messages that you might see on your blood pressure monitor and what you can do to try to resolve the problem. If you cannot resolve a problem, contact your care provider. Message Description EC02 The blood pressure monitor could not detect a stable pressure signal.
Message EC10 EC20 EC87 Description The blood pressure monitor was not able to connect to your telehealth appliance. Place your blood pressure monitor closer to your telehealth appliance, and try again The blood pressure monitor could not transmit your measurements to your telehealth appliance. Place your blood pressure monitor closer to your telehealth appliance, ensure that telehealth appliance is on, and then try again. The cuff did not inflate properly.
Message Description EC98 An error occurred with the internal software of the blood pressure monitor. Try again. 7. Symbols and Functional Features Front of the blood pressure monitor Symbol Description F.03D.601.424 Rev. A 12/2012 | Robert Bosch Healthcare Systems, Inc.
Please fill in the blank Symbol 15 | en Description Robert Bosch Healthcare Systems, Inc. | F.03D.601.424 Rev.
Top of the blood pressure monitor Symbol Description Battery compartment (bottom) Symbol Description F.03D.601.424 Rev. A 12/2012 | Robert Bosch Healthcare Systems, Inc.
Symbol Description 8. Cleaning, maintenance, and disposal To clean the blood pressure monitor, use a soft cloth dampened with water. Do not use any chemical or scouring agents. Ensure that liquid cannot penetrate into the blood pressure monitor. If your blood pressure monitor requires maintenance, contact your care provider. Contact your care provider regarding the disposal of your blood pressure monitor. Ensure that old batteries are disposed of in an approved location, separately from domestic waste.
9. Technical data Measurement range: 20 – 250 mmHG Protection class: IP21 Operating temperature: 41°F (5°C) to 104°F (40°C) Relative humidity: 15%%-93%%;%%Atmospheric pressure: 80kPa-105kPa%% Storage temperature: -13°F (-25°C) to 158°F (70°C) Relative humidity: 15%%-93%%;%%Atmospheric pressure: 80kPa-105kPa%% Power: 4 AA (R6P) batteries, 6V Cuff size: 8.6 in x 12.
10. Regulatory Intended Use: The Bosch Branded Blood Pressure Monitoring (BPM) System is intended for use by adults, to measure systolic and diastolic pressures as well as pulse rate, by the oscillometric method, in a telehealth system and for personal home use. Part 15 Warning: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Do not attempt to open or modify the device.
radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
Industry Canada Statement This device complies with Industry Canada licence‐exempt RSS‐210. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. This Class B digital apparatus complies with Canadian ICES‐003. FCC ID: NMTBP5000BT IC: 3277A‐BP5000BT Canadian RepreSentative Information Compamy Name: TJS Technical Services Inc.
I revise some cotent next page/ RoHS Compliance WEEE Compliance Bluetooth® Consult Instructions for Use Use the blood pressure monitor indoors only. Tested to Comply With FCC Standards Contains FCC ID: R47F2M03GX Keep dry. BF symbol for “applied parts” Keep away from heat. Manufacturer: Robert Bosch Healthcare Systems, Inc. Shipping temperature range. F.03D.601.424 Rev. A 12/2012 | Robert Bosch Healthcare Systems, Inc. 2118 Beechgrove Pl.
WEEE Compliance Consult Instructions for Use Use the blood pressure monitor indoors only. RoHS Compliance Bluetooth® EN 60950 Compliance “ Keep away from heat. Shipping temperature range. ” BF symbol for applied parts Robert Bosch Healthcare Systems, Inc. 2118 Beechgrove Pl.