User's Manual

IDS Ingegneria Dei Sistemi S.p.A. Protocol: MN/2009/030 - Rev. 1.2
Safe Rail System Data Acquisition Software
35
/
76
3.
The number of
Samples per scan
can be varied by choosing one of the
preset values available in the
Sample per scan
field (Fig. 4-5). However,
this value can’t be lower than the one indicated in the
Min req.
sample/scan
.
Fig. 4-5 – Samples per scan
4.
The electromagnetic wave propagation speed can then be set in the
Propagation speed
field (see Fig. 4-6). We recommend you accept the
default value unless you have specific information on the types of ground
present in the area to be investigated.
Fig. 4-6 – Propagation speed field
4.3 Setting up the Calibration
This paragraph describes the procedure used to calibrate the parameters used in
the real-time processing of radar data.
1.
Press the button shown in Fig. 4-2 to view the window
shown in Fig. 4-7, then perform the automatic calibration procedure by
moving the antennas across the surface to be scanned for a distance of at
least one meter.
2.
Stop the calibration phase by pressing the button in the
window shown in Fig. 4-7. This permits the system to save the filter
parameters required to view the radar maps in real time.