User Guide

13
L1L2 Measurement
Select one phase to be the common phase. (eg. S or L2)
Connect the black test probe to the common phase (S or L2)
and the red test probe to another phase (eg. R or L1).
Using the trigger to open the jaws, fully enclose the phase
(R or L1) within the clamp.
Release the trigger and be sure that the jaws have closed
completely.
Wait until the reading is stable, and press the READ NEXT
button to capture the reading, storing the data in memory.
The LCD display will show a flashing T to indicate that the
meter is ready for the next measurement.
Remove the clamp and test probes from the phases tested in
the first measurement.
L3L2 Measurement
Connect the black test probe to the common phase (S or L2)
and the red test probe to the third phase (eg. T or L3)
Using the trigger to open the jaws, fully enclose the phase (T
or L3) within the clamp.
Release the trigger and be sure that the jaws have closed
completely.
Wait until the reading is stable, and press the READ NEXT
button to capture the reading, storing the data in memory.
Calculations
14
Note: Once a phase is selected as common, this selection
cannot be changed. For example, if S or L2 phase is select-
ed, S is always connected to COM during measurement of
W
RS(L1L2)
in 3
φ
3W unbalanced power.
3ØP
F
A
C
+
D
C
2
0
0
0
K
W
2
0
0
0
A
OF
F
P
F
K
W
H
z
V
3
Ø3W
3Ø4W
KVARKVA
VA
D
C
A
/
W
Z
E
R
O
R
E
C
3
Ø
K
V
A
R
K
V
A
R
E
A
D
N
E
X
T
HOLD
C
A
T
I
I
I
-
6
0
0
V
Power Clamp
True RMS
6
1
-
8
0
2
3ØPF
AC+DC
2000KW
2000A
OFF
PFKW
HzV
3Ø3W
3Ø4W
KVARKVA
VA
DC
A/W
ZERO
REC
3
Ø
K
V
AR
KVA
READ
NE
X
T
HOLD
CAT III - 600V
P
o
w
e
r
C
l
a
m
p
True
R
M
S
6
1
-
8
0
2
3ØPF
AC+DC
2000KW
2000A
OFF
PFKW
HzV
3Ø3W
3Ø4W
KVARKVA
VA
DC
A/W
ZERO
REC
3
Ø
KV
AR
KVA
READ
NEXT
HOLD
CAT III - 600V
P
o
w
e
r
C
l
a
m
p
True
R
M
S
6
1
-
8
0
2