User Guide
14
Diode Testing
Function Resolution Accuracy Max. Test Max. Open
Current Circuit
Voltage
1mV +
(1.5% +5) 1.5mA 3V
* For 0.4V to 0.8V.
Diode Check:
1. Turn off power to the device or circuit that is being tested
and discharge all capacitors.
2. Plug the Black test lead into the COM port and the Red
test lead into the VΩHz port.
3. Set the rotary switch to the Ω position.
4. Press the blue button until is shown on the display
5. Connect the test leads to the diode. Normally the forward
voltage drop of a good silicone diode is shown between
400V and 0.9V. If the diode under test is defective, “000”
(short circuit) or “OL” (non-conductive) is displayed.
Overload Protection: 600V max
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2
Circuit Connection:
Capacitance
Range Resolution Max Display Accuracy
4nF 1pF 4.000 +
(3.0% +10)
40nF 10pF 40.00 +(3.0% +5)
400nF 100pF 400.0
4µF 1nF 4.000
40µF 10nF 40.00
400µF 100nF 400.0
4mF 1µF 4.000 +
(5.0% +20)*
40mF 10µF 40.00
* In these two ranges the reading may be rolling within specification
To Measure Capacitance:
1. Turn the power off to the circuit or device that is to be measured
and discharge all capacitors before attempting a measurement.
2. Plug the Black test lead into the COM port and the Red test lead
into the VΩHz port.
3. Set the rotary switch to the position.
4. Connect the leads to the circuit to be measured.
5. Measure Capacitance
In order to ensure the best accuracy in measurement of low capacitance,
open the test leads before measurement and
memorize the test probe capacitance. It is necessary to
subtract the capacitance of the test probes themselves, or use the rela-
tive mode to compensate for the test leads.
Overload Protection: 600V
13
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2