User Guide
12
Resistance (Ohms)
Range Resolution Max Display Accuracy
400Ω 0.1Ω 400.0 ±(0.7% +3)
4kΩ 1Ω 4.000 ±(0.4% +2)
40kΩ 10Ω 40.00
400kΩ 100Ω 400.0
4MΩ 1KΩ 4.000 ±(0.6% +3)
40MΩ 10KΩ 40.00 ±(1.5% +5)
To Measure Resistance:
1. Turn the power off to the circuit or device that is to be
measured and discharge all capacitors before attempting a
measurement.
2. Plug the Black test lead into the COM port and the Red test
lead into the VΩHz port.
3. Set the rotary switch to the Ω position.
4. For correct reading, ensure that the device being tested
contains no voltage.
5. Press the range button to select the proper range of the
meter.
Overload Protection: 600V
Open Circuit Voltage: -1.3V (approx.)
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2
Circuit Connection:
11
Frequency/RPM
Range Resolution Max Display Sensitivity Accuracy
4kHz/40kRPM 1Hz/30RPM 4.000/40.00 150mV Frequency:
≥2.0 Hz 0.01% +1
1.5V < 20 Hz RPM:
0.01% + 10
40kHz/400kRPM 10Hz/300RPM 40.00/400.0
400kHz/4MRPM 100Hz/3KRPM 400.0/4.000
4MHz/40MRPM 1kHz/30KRPM 4.000/40.00 300mV
40MHz/400MRPM 10kHz/300KRPM 40.00/400.0 1V
400MHz/4000MRPM 100kHz/3MRPM 400.0/4000 ***
*This spec is not guaranteed.
To Measure Frequency:
1. Plug the Black test lead into the COM port and the Red
test lead into the VΩHz port.
2. Set the rotary switch to the HZ RPM position.
3. For RPM, Push the blue button until RPM is shown on the
display.
4. Connect the meter in parallel with the load or circuit.
5. Measure Frequency or RPM.
Overload Protection: 600V
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2
Circuit Connection: