User Guide
DC Current
Range Resolution Max Display Accuracy Voltage
Burden
40mA 10µA 40.00 +
(0.4% +2) 300mV max
400mA 0.1mA 400.0 +
(0.5% +2)
3V max
10A 10mA 10.00 +
(1.0% +3) 3V max
To Measure DC Current:
1. Plug the Black test lead into the COM port and the Red
test lead into the mA or A port.
2. Set the rotary switch to the mA or A position.
3. Push the blue button until DC is shown on the display.
4. Connect the meter in series with the load or circuit.
5. Measure DC Voltage.
Overload Protection: 1A, 600V IR 10kA fuse for mA input
15A, 600V IR 100kA fuse for A input
10
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2
Circuit Connection:
AC Current
Range Resolution Max Display Accuracy Voltage
Burden
40mA 1µA 40.00 +
(1.0% +5) 300mV max
400mA 0.1mA 400.0 3V max
10A 10mA 10.00 +(2.0% +5) 3V max
To Measure AC Current:
1. Plug the Black test lead into the COM port and the Red
test lead into the mA or A port.
2. Set the rotary switch to the mA or A position.
3. Push the blue button until AC is shown on the display.
4. Connect the meter in series with the load or circuit.
5. Measure AC Voltage.
Frequency Response: 40Hz ~ 1kHz
Overload Protection: 1A, 600V IR 10kA fuse for mA input
15A, 600V IR 100kA fuse for A input
9
MIN/MAX
~Hz PEAK
HOLD
Data Acquisition
2
Circuit Connection: