User Guide
12
Resistance (Ohms)
Range Resolution MaxDisplay Accuracy
400Ω 0.1Ω 400.0 ±(0.7% +3)
4kΩ 1Ω 4.000 ±(0.4% +3)
40kΩ 10Ω 40.00
400kΩ 100Ω 400.0
4MΩ 1kΩ 4.000 ±(0.6%+3)
40MΩ 10kΩ 40.00 ±(1.5% +5)
Open Circuit Voltage: -1.3V approx.
To Measure Resistance:
1. Turn the power off to the circuit or device that is to be
measured and discharge all capacitors before attempting
a measurement.
2. Plug the black test lead into the COM port and the red test
lead into the VΩHz port.
3. Set the rotary switch to the Ω position.
4. For correct reading, ensure that the device being tested
contains no voltage.
5. Press the range button to select the proper range of the
meter.
6. Measure resistance.
11
Frequency with Hz Switch
Function Range Sensitivity Accuracy
AC Voltage 400mV 40mV rms 0.01%
4V 0.2V rms reading
40V 2V rms +5 digits
400V 20V rms
750V 200V rms
AC Current 40mA 8mA rms
400mA 80mA rms
10A 8A rms
To Measure Frequency with the Hz Switch
This switch is used to quickly view frequency during a mea-
surement of AC voltage or current.
1. Push ~ Hz during an AC voltage or current measurement.
2. The LCD will display the frequency measurement of the cir-
cuit.
3. Push ~Hz again to return to the AC voltage or current mea-
surement.
Hz
True RMS
Hz
True RMS
Circuit Connection:
˜
˜