#61-800 1 800 Series Power Clamp Meter 61-800 Power Clamp 2 HOLD OUTPUT 1mV/A 1200KW KW OFF 2500A 600V KW 250V 2100A V 3 4 V A A °F True RMS 5 61-800 DC A/W ZERO Power Clamp MIN MAX REL 7 V A A °F True RMS REL MIN MAX 6 CAT III - 600V HOLD OUTPUT 1mV/A 1200KW KW OFF 2500A 600V KW 250V 2100A V DC A/W ZERO ® RANGE ® 1. Transformer Jaw 2. Transformer Trigger 3. Data Hold Button 4. Rotary Function Switch 5. LCD Display 6. Function Buttons 7.
WARNING! 1. DO NOT UNDER ANY CIRCUMSTANCES EXCEED THESE RATINGS: - Voltage is not to exceed 800 Volts. - Wattage, Resistance and Continuity functions are not to be performed on circuits capable of delivering greater than 600 Volts. - Current measurements are not to be performed on circuits capable of delivering greater than 500 Volts 2. To avoid electrical shock hazards and/or damage to the meter: - Do not exceed the voltage ratings for the meter. Use caution when measuring voltage.
MIN/MAX Button • The Min/Max button is used to toggle between the minimum and maximum values captured by the clamp meter. • Press once to display the minimum value. • Press again to display the maximum value. • Press a third time to return to normal measurement mode. A 61-800 Range Button • Pressing the range button enables manual ranging. • While in manual range mode, pressing the range button toggles between the ranges in the selected function.
DC Current Measurements DC Current Range 400A 2000A 2000-2500A Resolution Accuracy 0.1A 1A 1A + ±(1.5% +3) Overload Protection: ±(2.0% +3) 3000A 4. Release the trigger and be sure that the jaws have closed completely. 5. If the meter is in auto-ranging mode, the range will be automatically selected. 6. If the meter is in manual ranging mode, press the range button to select the desired range. 7. Measure DC current.
To Connect to a Monitoring Device 1. The analog output of the meter is proportional to the current measurement (1mV/A). 2. Connect the monitoring device to the input terminals of the power clamp meter. The red terminal is positive and the black terminal is negative . 3. Connect these two terminals to an oscilloscope, chart recorder or data logger to observe the wave form, or record the data for future analysis.
AC+DC Watt Measurement Temperature Measurement AC+DC Watt Range (250V max) 40kW 400kW Resolution Accuracy 0.01kW 0.1kW ±(2.5% +5) AC+DC Watt Range (600V max) 400kW 1200kW Resolution Accuracy 0.1kW 1kW ±(2.5% +5) Overload Protection: 600VAC Temperature Range -50 to –400°F -400 to 1000°F Overload Protection: Resolution 0.
To Measure Resistance and Continuity 1. Set the rotary switch to . 2. Insert the black lead into the COM port and the red lead into the port. 3. Connect the test leads to the circuit to be measured 4. Measure resistance. 5. If the resistance is lower than 40Ω, the meter beeps. General Specifications Indoor Use Conductor Size: Cable 55mm (approx.) Bus Bar 65mm (D) x 24mm (W) Battery Type: 9V Display: 3-3/4 LCD with 40 seg.
Lifetime Limited Warranty This meter is warranted to the original purchaser against defects in material or workmanship for the lifetime of the meter. During this warranty period, IDEAL INDUSTRIES, INC. will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument.