Ranges & Accuracies:
AC Converter: 61-772, 61-774 models are true rms sensing.
Accuracy: Accuracy is specified as +/-(a percentage of the reading + a fixed amount) at
23°C±5°C (73.4°F ± 9°F), less than 75% relative humidity.
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F)
*Accuracy stated for crest factor 3 at full scale and 6 at half scale
Input impedance: 400mV: >50M; 4V: 10M; 40V~ 1000V: 9.1M
Accuracy
Function
Range and Overload
Resolution
61-772 61-774
Protection
400.0/1000A
(0~600A, 50-60Hz) 1.7% + 10
400.0/1000A (0~600A, 60-400Hz) 3.0% + 10
AC Current*
400.0/1000A (600~1000A, 50-60Hz) 2.5% + 10
1000 AAC
400.0/1000A (600~1000A, 60-400Hz) 3.5% + 10
400.0/1000A
(0~400A) N/A 1.5% + 5
DC Current 400.0/1000A
(400~800A)
N/A 2.0% + 5 1000 ADC
400.0/1000A
(800~1000A) N/A 3.0% + 5
400.0m (50-60Hz) 1.2% + 8
1000VDC or
AC Voltage 4.000/40.00/400.0V
(50-500Hz) 1.2% + 8
750V (
50-500Hz) 1.5% + 8
750VAC rms
DC Voltage 400.0m/4.000/40.00/400.0/1000V 0.5% + 2 1000VDC or 750 VAC rms
400.0/4.000k/40.00k/400.0k 1.0% + 4
Resistance 4.000M 5.0% + 4 600 VDC or AC rms
40.00M (specified to 10M) 12.0% + 5
4.000µF N/A 3.0% + 10
Capacitance 40.00µ/400.0µ F N/A 3.0% + 5 600 VDC or AC rms
4.000mF N/A 5.0% + 20
Audible indication < 35 ••
Continuity
Response time: 500ms
600 VDC or AC rms