Insulation Testing
1. Turn the instrument ‘On‘ by selecting the MΩ range.
2. Connect the test probes to the isolated circuit under test.
3. Press and hold the push button; the display shows the insulation value.The
reading will remain displayed for a few seconds after the push button is
released. As an additional safety feature, instruments with
1000V range will
flash symbol before performing a test.
4. Release the push button before removing the test leads (to enable the
instrument to discharge the circuit under test).
5. On completion switch to the ‘
OFF’ position. Alternatively auto shut off
operates after 12 minutes of inactivity.
SPECIFICATION
Insulation
Test Voltage Accuracy: -0% + 30% (over full operating temperature)
into 0 to 1mA load
Measuring Range: 0,01MΩ - 999MΩ (Digital)
0,1MΩ - ∞ (Analogue)
Short Cct. Current: Less than 2mA
Accuracy (at 20°C): ± 3% ±2 digits up to 10MΩ
± 5% ±2 digits up to 100MΩ
± 30% up to 999MΩ
Output Noise Voltage: Typically 2V pk to pk at 20kHz (at 1mA load)
Hum Rejection: <10% error with 100µA RMS (0,2MΩ to
∞
)
Continuity
Measuring Range: 0,01Ω - 99,9Ω
Open Cct. Voltage: 5V ±1V
Accuracy (at 20°C): 0,01Ω - 9,99Ω ±3% ± 2 digits (I
SC
>200mA)
10
Ω − 100Ω ± 5% ± 2 digits (I
SC
>20mA)
Lead res. comp: 0 to 9,99Ω
Hum Rejection: <3% error with 1V RMS (0,2Ω to 50Ω)
Continuity Beeper: On position, the beeper sounds at <5Ω resistance
Environmental Conditions
Temperature Coefficient: <0,1% per °C
Temperature Range:
Operating: -20°C to +40°C (full measurement range)
-20°C to +60°C (to 100 MΩ maximum)
Storage: -25°C to +65°C
Humidity: 90% RH at 40°C max
General Specifications
Display: 3 digit L.C.D. Maximum reading 999
Auto Shut off: Operates after 5 minutes of inactivity by the
instrument in Continuity position and 12 minutes in Insulation position. Beep
notes pre-empt auto shut off. The instrument can be switched back on by
pressing the test button, or by selecting ‘
OFF‘ and then the required position.
Default Live Circuit Warning
When more than 25 V is applied to the terminals, the instruments default to a
1000V
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
0,01MΩ
0,1MΩ
1,0MΩ
10MΩ
100MΩ
Load
Terminal Voltage
1000 V Range
500 V Range
0.5MΩ
250 V Range
0,2MΩ
0,5MΩ 1MΩ
2MΩ
5MΩ
0,2MΩ
0,5MΩ
1MΩ
2MΩ
5MΩ
10MΩ
Minimum Indicated Value
Terminal Characteristics
Limiting Value