User Guide
19
Capacitance Measurements
Range Resolution Accuracy Accuracy
(61-633) (61-635)
4nF 1pF ±(1.9% +20) ±(0.9% +20)
40nF 10pF
400nF 100pF
4µF 1nF
40µF 10nF ±(2.9% +20) ±(1.9% +20)
400µF 100nF
4mF 1µF ±(3.9% +20) ±(2.9% +20)
10mF 10µF
To Measure Capacitance:
1. Turn the power off to the cir-
cuit or device that is to be
measured and discharge all capacitors before attempting a
measurement.
2. Plug the Black test lead into the COM port and the Red test
lead into the VΩ port.
3. Set the rotary switch to the µ position.
ND 2352-1 61-633 630 Series Ins 12/22/04 9:49 AM Page 19