User Guide
Page 6 
CALIBRATION 
(1) Calibration Preparation 
(1.1) This procedure consists of the following sections. 
 (1) Calibration Preparation 
 (2-6) Adjustments 
 (7) Disassembly and Component Layout 
 (8) Trouble-shooting 
 (9) List of spare parts 
(1.2) Adjustment procedure will require the following tools and fixed standards 
 Prepare following items for adjustment. 
 Oscilloscope, 20 MHz or better. See Figure 1 
 General resistors @ .5% or better (0.1/10/20/200/1000), 
 Wire of fixture to form a loop circuit with each of the resistors as part of the loop. See Figure 2 
 Calibrated current source .1mA to 30A, with an accuracy of 0.5% or better. 
 Figure 1: Oscilloscope to check waveforms 
(1.3) Connect the probe and alligator clip of the oscilloscope as shown in figure one to form a circle (Short the 
input to common or ground of the scope). 
(1.4) Connect a wire and each general resistor (0.1/10/20/200/1000) like a circle respectively. 
Construct the loop using low 
resistive wire. Minimize 
connectivity resistance 
Figure 2 
(1.5) Open and close the jaws of 61-920 several times, and confirm that the jaws close by themselves. 
(1.6) Use care in removing front panel decal. The adjustment locations shown in Figure 3 are located 
underneath the decal. Note: plastic switch extensions will be loose with the decal removed. 
Form number TM61920  Rev 4 July 2008 