User Guide
Page 11
(8). Trouble shooting (cont’d.)   
  Symptoms Possible Causes  Remedies 
Insufficient capacity of battery  Replace the battery if “BATT” mark is 
displayed on the LCD. 
Bad fit of Jaw.  Make sure that the Jaw is completely closed. 
Replace the jaw if it cannot be completely 
closed and do re-adjustment. 
Re-adjustment of PCB  Check above points first. If there is no 
problem, PCB needs to be re-adjusted. 
When measuring 
resistance, 
error of measurement is 
notable. 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
Insufficient capacity of battery  Replace the battery if “BATT” mark is 
displayed on the LCD. 
Bad fit of jaw.  Make sure that the jaw is completely closed. 
Replace the jaw if it cannot be completely 
closed and do re-adjustment. 
Re-adjustment of PCB  Check above points first. If there is no 
problem, PCB needs to be re-adjusted. 
When measuring 
AC current, error of 
measurement is notable. 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
Button spacer is missing.  Put the button spacer at the specific 
location. 
Any button doesn’t work. 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
Poor contact of buzzer  Soldered point at buzzer should be checked. Buzzer doesn’t sound. 
(when a button is operated 
or in Buzzer mode) 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
Wrong function is selected.  In case that “” isn’t displayed on the LCD 
as a unit, press A/ button several times and 
select the resistance measurement function. 
“OPEN” mark doesn’t 
appear while jaws are 
opend. 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
A break in jaw wiring  Make a continuity test of jaw wiring. If there 
is no continuity, replace the jaw and do 
re-adjustment. 
“OPEN” mark doesn’t go 
off after closing the jaws. 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 
Measuring the current flowing 
line. 
Turn off the device, which is generating the 
current, to minimize the current. 
Two or more 61-920 are clamping 
onto one same line 
Use only one 61-920 per line. 
“NOISE” mark doesn’t 
go off 
Defect of circuit component  Check above points first. If there is no 
problem, replace the PCB, and do 
re-adjustment. 