Use and Care Guide
Page 5 
Table 2 Calibration Range Inputs: Model 61-795 
Range  Function  Input the calibrator standard value 
1 DCV  DC600.0V 
2 ACV  AC600.0V 
3 
Ω
190.0Ω 
4 
Ω
0.00Ω 
5 
LOΩ 19.00Ω 
6 
LOΩ 0.00Ω 
7 
LOΩ 5.00Ω 
8 
LOΩ 0.00Ω 
9 
250V/MΩ 1000MΩ 
10 
250V/MΩ 100.0MΩ 
11 
250V/MΩ 10.00MΩ 
12 
250V/MΩ 1.000MΩ 
13 
250V/MΩ 0.000MΩ 
14 
500V/MΩ 1000MΩ 
15 
500V/MΩ 100.0MΩ 
16 
500V/MΩ 10.00MΩ 
16 
500V/MΩ 1.000MΩ 
18 
500V/MΩ 0.000MΩ 
19 
1000V/MΩ 1000MΩ 
20 
1000V/MΩ 100.0MΩ 
21 
1000V/MΩ 10.00MΩ 
22 
1000V/MΩ 1.000MΩ 
23 
1000V/MΩ 0.000MΩ 
Battery Replacement (Refer to Figure 2) 
1. Disconnect the test leads from any circuit under test. 
2. Press the 
 key to switch off the instrument. 
3. Remove the hood and battery compartment cover by using a screwdriver. 
4. Remove the batteries replacing them with new ones all of the same type (1.5×6 NEDA 15F IEC R6 JIS 
UM-3) (Alkaline batteries are recommended.) making sure of polarity while installing. 
5. Install bottom case cover and secure with screws. 
6. Install the hood. 
Replacing Fuse (Refer to Figure 2) 
1. Disconnect the test leads from any circuit under test. 
2. Press the 
 key to switch off the instrument. 
3. Remove the hood and battery compartment cover by using a screwdriver. 
4. Loosen the screws from the case bottom. Remove the case bottom. 
5. Remove input cover using a screwdriver. 
6. Check the fuse with a digital multimeter with a source current < 10mA in low resistance range. 
 (Replace fuse with a BBS 1A/600V, 10.3*38 (1.5x0.41 inch) UL/CSA, interrupting rating 10kA.) 
7. Install input cover and secure with screw. 
8. Replace the case bottom and secure with screw. 
9. Replace the battery compartment cover, pressing to close it, and secure with screw. 
10. Install the hood. 
Form number TM61795  Rev 3 Jun 2008 