Use and Care Guide
Page 3 
PERFORMANCE VERIFICATIONS 
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning 
correctly. If the meter does not conform to any of the listed limits, the calibration procedure must be 
performed. 
Check the fuse as a BBS, 1A/600V, 10.4*35 (0.41x1.38 inch) UL/CSA, interrupting rating 10kA. 
Check case, leads, display, etc, for any defects. 
Table 1 Performance Verification: Model 61-795 
Function Setting 
/Range 
Apply Reading  Specification 
DCV  550V DC  547.2 to 552.8  [±0.5% ±1 digits] 
ACV  550V AC  545.2 to 554.8  [±0.8% ±4 digits] 
Continuity Ω/  
<30Ω  Buzzer sounds  Not specified 
Resistance:     
LOΩ  Simulated leads  .08 to .12  [±0.02Ω + 0.05 per lead] 
LOΩ 0Ω  -0.03 to +0.03  [±0.03Ω] 
LOΩ 5Ω  4.87 to 5.13  [±2.0% ±3 digits] 
LOΩ 19Ω  18.59 to 19.41  [±2.0% ±3 digits] 
Ω/  
190Ω  185.9 to 194.1  [±2.0% ±3 digits] 
LOΩ 10Ω  20.2mA to 25.8mA  Source check 
LOΩ 1Ω  202mA to 258mA  Source check 
Insulation Resistance:     
MΩ, 250V  10MΩ  250 to 275Vdc  Source Check 
MΩ, 500V  10MΩ  500 to 550Vdc  Source Check 
MΩ 1000V  10MΩ  1000 to 1100Vdc  Source Check 
0Ω  0.001 to 0.010  ±10 digits 
100kΩ  0.090 to 0.110  ±10 digits 
1MΩ  0.975 to 1.025  [±2.0% ±5 digits] 
10MΩ  9.75 to 10.25  [±2.0% ±5 digits] 
MΩ, 250V 
100MΩ  94.5 to 105.5  [±5.0% ±5 digits] 
100kΩ  0.085 to 0.115  ±15 digits 
1MΩ  0.975 to 1.025  [±2.0% ±5 digits] 
10MΩ  9.75 to 10.25  [±2.0% ±5 digits] 
100MΩ  97.5 to 102.5  [±2.0% ±5 digits] 
MΩ, 500V 
1000MΩ  945 to 1055  [±5.0% ±5 digits] 
100kΩ  0.085 to 0.015  ±15 digits 
1MΩ  0.975 to 1.025  [±2.0% ±5 digits] 
10MΩ  9.75 to 10.25  [±2.0% ±5 digits] 
100MΩ  97.5 to 102.5  [±2.0% ±5 digits] 
MΩ, 1000V 
1000MΩ  965 to 1035  [±3.0% ±5 digits] 
MΩ, 500V  500KΩ <500V  Source Check 
Check on each 
Voltage range 
0MΩ <3.0mA  Source Check 
Form number TM61795  Rev 3 Jun 2008 