Electrical Specifications
Function Range Resolution Accuracy
61-557 ±(a%+b)
AC Voltage (V) 6-600V 1V ±(3.0%+2)
LoZ ACV (V) 6-600V 1V ±(3.0%+2)
DC Voltage (V) 6-600V 1V ±(3.0%+2)
GFCI (mA) 6mA~9mA 102V,>6.0mA ±(2.0%+5)
1. Overload Protection: 720VRMS
2. Accuracy a is % of reading and b is LSD (Least Significant Digit).
17
Function Range Resolution Accuracy
61-557 ±(a%+b)
Continuity 1.0 Ω
10kΩ : Buzzer
beeps and red
indicator LED
illuminates
continuously
15kΩ : No buzzer
beep
NCV 90-600V
90V/(50 to 60Hz), with direct wire
contact, red indicator LED flashes at a
frequency of 3Hz , and the buzzer beeps
continuously
Function Description
LCD 600 count LCD. Displays a “-” symbol for all negative readings, displays
“AC” for alternating current or “DC” for direct current.
LCD Backlight White backlight.
Flashlight White flashlight works independently of LCD backlight.
High Voltage
Alarm
1) Effective setting: ACV / DCV.
2) In ACV/DCV, voltage measurement value: when measuring >30V AC or
30V DC, LCD displays high voltage and PPE. The Red LED illuminates, a
PPE icon will appear in the screen and the buzzer will beep for one second.
Note: For models with SN earlier than 2110XXXX, only the PPE
icon will appear. The Red LED will not illuminate and there is
NO beep.
3) If 600V AC/DC is exceeded, the LCD displays “OL”, the buzzer beeps
On/Off continuously and the high voltage alarm (PPE symbol) is displayed.
Over Range
Indication
LCD displays “OL” when over range is encountered.
Low Battery
Indication
The “ ” symbol appears in the LCD display with insufficient
voltage. Then the buzzer beeps for 5 seconds and the unit powers down.
This ensures accurate readings.
APO The meter automatically beeps 5 times and powers down after 15 minutes
of detecting no activity in any of the three following modes; Voltage Test,
NCVT, or Continuity.
Environmental Specifications
Operating Temperature: 32ºF to 104ºF (0ºC to 40ºC) (<80%RH)
Operating Altitude:
6500 ft (2000 m)
Storage Temperature:
14ºF to 122ºF (-10ºC to 50ºC) (<80%RH)
Intended for indoor use.