Instructions / Assembly

Page 18
Testing the Resistance Function (for 61-340 / 61-342)
To verify the accuracy of the resistance function, do the following:
1. Connect the calibrator to
and COM on the meter.
position. Press the SELECT button to select the function. 2. Turn the rotary switch to the
3. Apply the inputs for step 1 - 7 in Table 5.
4. Compare the meter display readings to the display readings in Table 5.
5. If the display reading falls outside of the range shown in Table 5, the meter does not meet
specification.
Table 5 Resistance Test:
Step Range Source Reading
1 400* 4.0 3.5 to 4.5
2 400* 300.0 295.9 to 304.1
3 4k 3.000K 2.968 to 3.032
4 40k 30.00K 29.68 to 30.32
5 400k 300.0K 296.8 to 303.2
6 4M 3.000M 2.962 to 3.038
7 40M 30.00M 29.35 to 30.65
*Lead resistance on the 400 range is not included in error.
Testing the Capacitance Function (for 61-340 / 61-342)
The meter measures capacitance by charging the capacitor with a known direct current, measuring the
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an
impedance bridge, a different reading may result. This variance is likely to be greater at higher
frequencies.
To verify the accuracy of the capacitance measuring function, do the following:
1. Apply the capacitor to the
and COM inputs on the meter. For steps 1 - 7 in Table 6.
2. Turn the rotary switch to the
position. Press the SELECT button to select the function.
Press the REL button to make sure the display is zero.
3. Apply the inputs for step 1 - 7 in Table 6.
Note: The meter selects the proper range automatically. Each measurement takes about one
second per range, readings >40.00μF will take 5 to 30 seconds.
4. Compare the meter display readings to the display readings in Table 6.
5. If the display reading falls outside of the range shown in Table 6, the meter does not meet
specification.
Form Number TM61340-2 Rev 3 January 2008