Table 6 Capacitance Test:
Step
Range
Source
Reading
1
40nF
10.00nF
9.60 to 10.40
2
40nF
30.00nF
29.00 to 31.00
3
400nF
300.0nF
290.5 to 309.5
4
F
3.000µF
2.905 to 3.095
5
40.00µF
30.00µF
29.05 to 30.95
6
400.0µF
300.0µF
239.5 to 360.5
7
4000µF
3000µF
2395 to 3605
Checking the Diode Test Function (for 61-340 / 61-342)
To check the diode test function, do the following:
1. Connect the test leads to the
and COM inputs on the meter.
2. Turn the meters rotary switch to the
position. Press the SELECT button to select
the
function. The display should display OL.
3. Insert a Si diode with correct polarity “+” and-”. The meter display should read within
0.5~0.7Vdc.
4. Reverse polarity on the Si diode. The meter display should read OL.
.
Checking the Continuity Function (for 61-340 / 61-342)
To check the diode test function, do the following:
1. Connect the test leads to the
and COM inputs on the meter.
2. Turn the meter’s rotary switch to the
position. Press the SELECT button to select the function.
The display should display OL.
3. Insert a 20 resistor. The meter’s internal beeper will sound.