Full Product Manual
Document: September1, 2009 Page 6of 10
Steps Load Measurement Current
1 No-Load V= No load Voltage H-N
N
/A
2
Load H-N, V1= Voltage H-N
3
Load H-N, V2 =Voltage H-G
4
Load H-G, V3=Voltage H-N
5 Load H-G, V4=Voltage H-G
To calculate the impedance of each conduct enter the values from the table above and
calculate the impedance each conductor
1
If the method used to measure current employs a shunt resistor the internal shunt
resistance of the meter should be taken into account in your calculations. On most
DMM’s this shunt resistance is around 0.1 ohm.
Enter “Calculated Impedance” measurements into Table 3 below.
Measure the PEAK voltage and enter in Table3
Measure Frequency and enter into Table 3
Measure Neutral to Ground Voltage and enter into Table 3
Enter calculated %VD for 12 amps, 15 amps, and 20 amps and enter in to Table 1 below
Reference Measured Voltage drop section
For best accuracy monitor RMS Voltage, Peak Voltage, N-G voltage during UUT test.
These measured values could change.
Performance Verification Preparation
To complete the performance verification the Measured Values column must be filled out
in Table 3 through standardization practice or calculated methods. See
STANDRADIZING A BRANCH CIRCUIT above. Then plug in the Surtest and take
readings to fill out the SureTest Reading column in Table 3
Note: The SureTest™ only takes reading when first plugged in. AC line voltage, peak
and Frequency are the only continuous updated reading. Ground impedance is only
measured when the meter goes to that reading. If you are on a GFI protected circuit the
ground impedance measurement will most likely trip the device.
Calculated Impedance
rZ
1
=(V-V
1
)/I
rN=(V
2
-V
1
)/I
rG=(V
3
-V
4
)/1
rH
1
=(rZ-rN)
1
rZ = Total branch circuit Impedance
rN = Impedance of neutral conductor
rG = Impedance of the Ground wire
rH= Impedance of the Hot or High
conductor