User Manual

Form Number TM61096 Rev 2 Nov 2010
Measurement Characteristics
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 80% R.H.
DC / AC Volts
Function
Range
Accuracy
V
1.3V ~ 750.0V
±(1.5% reading + 3 digits)
50Hz ~ 500Hz
V
1.8V ~ 1000V
±(1% reading + 2 digits)
-0.6V ~ -1000V
±(1% reading + 4 digits)
Overload protection: 750Vrms // 1000VDC
Max. Operation time: DT=30s for ≥ 30V
Input Impedance: ≥ 4K for input voltage up to 30V. Impedance increase with input voltage to 330KΩ at
1000V.
Resistance & Continuity
Function
Range
Accuracy
Ω
0.0Ω ~ 99.9Ω
±(2% reading + 1Ω)
100Ω ~ 2000Ω
±(2% reading + 2 digits)
Overload protection: 750Vrms // 1000VDC
Max. Open Voltage: 1.5V
Continuity check: Internal sounds activates if the resistance of the circuit under test is less than 25Ω.
It will then turn off if the resistance is increased beyond 400Ω.
Operating temperature: 0°C ~ 40°C
AC Current
Function
Range
Accuracy
A
0.6A ~ 200.0A
±(1.8% reading + 3 digits)*
50Hz ~ 60Hz
Overload protection: 400A
* ±(1.8% reading + 6 digits) ≤ 50A
4