99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL 61-096 Split Jaw Smart Meter The Service Information provides the following information : • Precautions and safety information • Specifications • Basic maintenance (cleaning, replacing the battery and fuses) • Performance test procedures • Calibration and calibration adjustment procedures Form Number: Revision: 2.
TABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 Specifications 3 General Specification 3 Voltage Specification 4 Resistance & Continuity Specification 4 Current Specification 4 Physical and environment characteristics 5 Certification and compliance 5 Required Equipment 6 Basic Maintenance 7 Opening the Meter Case 7 Replacing the Battery 7 Performance Tests 8 Testing the Voltage Function 8 Testing the Resistance Function 9 Test
Introduction Warning To avoid shock or injury, do not perform the verification tests or calibration procedures described in the manual unless you are qualified to do so. The information provided in this document is for the use of qualified personnel only. Caution The 61-096 contain parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. Precautions and Safety Information Use the Meter only as described in the Service Manual.
SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or products connected to it. To avoid potential hazards, use the product only as specified. CAUTION: These statements identify conditions or practices that could result in damage to the equipment or other property. WARNING: These statements identify conditions or practices that could result in personal injury or loss of life. Specific precautions Do not operate without covers.
SPECIFICATIONS All specifications are warranted unless noted typical and apply to the 61-096 Stated accuracies are at 23°C±5°C at less than 80% relative humidity and without the battery indicator displayed.
Measurement Characteristics Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 80% R.H. DC / AC Volts Function V V Range 1.3V ~ 750.0V Accuracy ±(1.5% reading + 3 digits) 50Hz ~ 500Hz 1.8V ~ 1000V ±(1% reading + 2 digits) -0.6V ~ -1000V ±(1% reading + 4 digits) Overload protection: 750Vrms // 1000VDC Max. Operation time: DT=30s for ≥ 30V Input Impedance: ≥ 4K for input voltage up to 30V. Impedance increase with input voltage to 330KΩ at 1000V.
Physical and Environmental Characteristics Characteristics Description Dimensions (W×H×L) 84mm×188mm×41mm Weight (with battery) 230g Environmental characteristics Description Temperature operating 0 to +50°C Non-Operating -20 to +60°C Humidity (operating) <80% R.H. Altitude Operating 2,000M (6560 ft.) Non-Operating 12,300M (40354 ft.) MIL-T-28800E TYPE II Class 5 Vibration & shock Operating 2.
Required Equipment Required equipment is listed in Table B. If the recommended models are not available, equipment with equivalent specifications may be used. Repairs or servicing should be performed only by qualified personnel. Table B. Required Equipment Equipment Calibrator Required Characteristics AC Voltage Range: 0 ~ 750V AC Accuracy: ±0.07% (Basic) Frequency Range: 40 ~ 1KHz Accuracy: ±2% DC Voltage Range: 0 ~ 1000V DC Accuracy: ±0.006% (Basic) Current Range: 0 ~ 10A Accuracy: AC (40Hz to 1KHz): ±0.
Basic Maintenance Warning To avoid shock, remove the test leads and any input signals before opening the case or replacing the battery. Opening the Meter Case Caution To avoid unintentional shock circuit, always place the uncovered Meter assembly on a protective surface. When the case of the Meter is open, circuit connections are exposed. To open the Meter case, and do the following: 1. Disconnect test leads from any live source, push the button to OFF, and remove the test leads from the front terminals.
Performance Tests The following performance tests verify the complete operability of the Meter and check the accuracy of each Meter function against the Meter’s specifications. Accuracy specifications are valid for a period of one year after calibration, when measured at an operating temperature of 18°C to 28°C and a maximum of 80% relative humidity.
6. Set the calibration for the voltage from step 1 to 8 in Table 2. 7. Compare the reading on the Meter display with the display reading shown in Table 2. 8. If the display reading falls outside of the range shown in Table 2, the meter does not meet specification. Table 2 DC Voltage Test: Step Input Reading 1 3.0V 2.8 to 3.2 2 -3.0V -2.6 to -3.4 3 30.0V 29.5 to 30.5 4 -30.0V -29.3 to -30.7 5 600.0V 593.8 to 606.2 6 -600.0V -593.6 to -606.4 7 900.0V 890.8 to 909.2 8 -900.0V -890.
Testing the AC Current Function To verify the accuracy of AC current measurement functions, do the following: 1. Open jaw of the meter around the suitable wire or conductor. 2. Push the button to start your meter. 3. Apply the inputs for steps 1-4 in Table 4. 4. For each input, compare the readings on the Meter display to the reading in Table 4. 5. If the display reading falls outside of the range shown in the Table 4, the meter does not meet specification.
CALIBRATION PROCEDURE Recalibrate your meter: It is recommended that the meter may be calibrated once year. Use the following procedure to calibrate the clamp meter. 1. Perform calibration at an ambient temperature and relative humidity (23°C±2°C and R.H.≤80%). Allow instrument to sit at this temperature for at least thirty minutes. 2. Disconnect the test leads from any circuit under test and turn off meter. 3.
VR13 VR14 VR12 VR15 VR17 VR16 Figure 1 61-096 Calibration Adjustment Points 12 Form Number TM61096 Rev 2 Nov 2010
VR1 Figure 2 61-096 Calibration Adjustment Points 13 Form Number TM61096 Rev 2 Nov 2010