User Manual
www.ideal-tek.com
Switzerland
8
System set up
Example of system setup when TEK-SCOPE PLUS Software is used:
DO NOT CONNECT THE DEVICE BOTH TO A MONITOR AND PC TOGETHER. 
If the computer and the monitor are used together, like extended or dual, the algorithm for magnication/
measuring will be effected.
With the following screens sizes: 20”(1600x900); 22”(1920x1080); 22”(1680x1050); 24”(1920x1080); 24”(1920x1200) 
measuring will have a tolerance of 1-2% without calibration.
Measuring
Put your object under the camera. Choose the magnication by the software buttons. 
Notice: using the remote control, the Zoom factor will be not updated and measuring will be wrong.
DO NOT USE REMOT CONTROL WHEN CONNECTED TO THE PC BUT USE THE SOFTWARE TOOLBAR. 
Press the draw button. Press LB (Left button) on your mouse, select the horizontal or vertical button and click 
the starting point. Then go to the second measuring point and press LB on your mouse. The measure value will 
appear on the screen.
Calibration
If you need a more accurate measuring, the calibration 
should be done. 
Put a ruler horizontally on the subject you wish to measure. 
Verify that your magnication is the one you want, select 
the drawing mode, and  select “calibration”  button. Draw 
an horizontal line and set your “own value” in mm, then 
press “set”.
Zoom value / rotation: how much the zoom moves vs. the scroll wheel
Editing Colors: change the color of text and drawings in the Draw Mode