Operation Manual

Chapter 5
Autotest
51
4. Connect Display handset (DH) and Remote handset (RH) with patch cord for
Display handset (DH), as shown on the display. Remember which connection of the
patch cord is connected to the Display handset (DH).
Illustration 3 5.9.1. Patch Cord (DH)
5. Use soft key to continue calibration. The Remote handset (RH) indicates
calibration, and a progress bar shows the status.
6. Perform the same patch cord connections to Display handset (DH) and Remote
handset (RH) as during the previous test steps.
Illustration 5.9. Open Patch Cord
7. On Display handset (DH) use soft key to start the last step of calibration.
8. On Remote handset (RH) use Autotest to start the last step of calibration.
Illustration 4 5.9.1. Calibration Complete
9. On Display handset (DH) press Escape to return to screen. The Remote handset
(RH) briefly displays that calibration was successful, and then shows ready state.